| Model Number |
97022HS |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer alleging discrepant high tni with triage cardiac lot t15228rn vs er method (hstni) for 1 patient.Symptoms: unknown diagnosis: unknown triage device thawed properly.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation conclusion: the customer's complaint was not replicated during in-house testing with retained devices of triage cardiac lot t14511rn.Retains of the complaint lot performed properly when tested with "normal" (apparently healthy) donors.No issues with tni recovery were observed.Manufacturing batch records for the lot were reviewed and found that the lot met release specifications.Based on the information available, there is no indication of a product deficiency or no corrective action is required.
|
| |
|