U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number 3543001
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 02/19/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
D4: udi: as the serial number for the device involved in the event was not provided, the full udi is currently not available.Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: material rupture mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient underwent breast augmentation revision with two 300cc mentor memorygel breast implants.Post-operatively, the patient was diagnosed, via physical examination, with left breast implant rupture and right breast capsular contracture (baker grade iii).As a result, the patient underwent bilateral breast implant removal and replacement surgery on (b)(6) 2025.The replacement devices were mentor gel implants.This medwatch form is for the left breast prosthesis.
 
Additional Manufacturer Narrative
On may 19, 2025, an analysis of the suspect medical device¿s photo was completed.Investigation summary: upon visual evaluation of the image provided in the complaint, the implant was observed ruptured.As the product involved in this complaint was not received, the device could not be analyzed according to our procedures.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.No corrective and preventive action (capa) is required now.
 
Additional Manufacturer Narrative
On june 30, 2025, the mentor failure analysis lab received the device for evaluation.The patient's date of birth and identifier were also provided and have populated accordingly.On july 4, 2025, the product investigation was completed.Device investigation summary: the product was returned to mentor for evaluation.Mentor conducted visual inspection of the returned device.Visual analysis of the returned device revealed that the breast implant was found to have a tear on the posterior view, measuring approximately 0.4 cm.In addition, siltex cracking was noted on the edge of the rupture.The evaluation determined that the possible cause of the rupture is consistent with normal wear.Siltex cracking is defined as a material separation occurring in the siltex layer and can eventually progress through the shell of siltex devices.Siltex cracking is ultimately a result of high stresses.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria, and that normal wear may have caused the observed defect.No corrective and preventive action (capa) is required now.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key22032208
Report Number1645337-2025-05356
Device Sequence Number18319444
Product Code FTR
UDI-Device Identifier00081317000860
UDI-Public(01)00081317000860(17)171031(10)6643363
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2017
Device Catalogue Number3543001
Device Lot Number6643363
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/22/2025
Supplement Date Received by Manufacturer05/19/2025
06/30/2025
Initial Report FDA Received Date05/15/2025
Supplement Report FDA Received Date06/11/2025
07/09/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/11/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-