| Model Number |
MA60AC |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A physician reported that during an intraocular lens implant procedure the haptic was broken during the implantation.There was no patient contact and no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in d.9., d.10., h.3., h.6.And h.11.The lens was returned for evaluation.Solution is dried on the lens.One haptic is broken in the distal area (not returned).The opposite haptic is bent in the gusset area.Just above the bent area the haptic is fractured.The tip of the bent haptic appears to have made contact with a heat source causing it to have a round swollen appearance.The optic is crushed over two posts in the lens case with the posts protruding through the optic.Loose lens material from the optic is around the posts.Product history records were reviewed and the documentation indicated the product met release criteria.The file indicates the use of a qualified cartridge and handpiece with a non-qualified viscoelastic.Broken haptic damage was observed.The root cause for the reported complaint may be related to a failure to follow the instructions for use (ifu).The viscoelastic indicated is not qualified for the lens company combination used.Company foldable iols are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs use holding forceps to grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd filled cartridge.The lens should be inserted until it is centered with or slightly past the outline etched into the top of the cartridge.Lenses with a 6.5 mm optic diameter may need to be pre-folded for easy entry into the back of the cartridge.Lens may be pre-folded by gently applying pressure to the edge of the lens while holding the central optic with forceps the trailing haptic will extend from the proximal end of the cartridge.Position the trailing haptic to the left of the haptic post as shown in the detailed view.This will allow the plunger to go past the haptic during the initial advancement of the plunger into the cartridge.Verify the lens is positioned on the bottom surface of the cartridge.The haptic should be maintained to the left of the post when the lens is loaded correctly.Accurate positioning of the lens will decrease the potential for optic and haptic damage.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|