• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOL-MILLENNIUM MEDICAL INC., CHICAGO, IL, USA. SOL-MILLENNIUM; MONOJECT BLUNT FILL NEEDLE 18G*1 1/2"

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOL-MILLENNIUM MEDICAL INC., CHICAGO, IL, USA. SOL-MILLENNIUM; MONOJECT BLUNT FILL NEEDLE 18G*1 1/2" Back to Search Results
Model Number 11811022
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There is no evidence of quality issues associated with this complaint based on the archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 04/06/2025.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
 
Event or Problem Description
" the customer reported that when using the cardinal needle to draw up medication, coring occurred resulting in stopper fragments from medication vial top appearing in syringe.
 
Additional Manufacturer Narrative
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.This supplemental mdr is being submitted to correct all manufacturer contact office information and date of event provided in the initial mdr [3014312726-2025-00075].The previously reported was incorrect.The correct manufacture contact information is below.(b)(4).The correct date of event is 01-april-2025.The correct product description is monoject blunt fill needle 18g*1 1/2".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL-MILLENNIUM
Common Device Name
MONOJECT BLUNT FILL NEEDLE 18G*1 1/2"
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC., CHICAGO, IL, USA.
311 s wacker dr, suite 4100
chicago IL 60606
Manufacturer (Section G)
ZHEJIANG KINDLY MEDICAL DEVICES CO., LTD
no.758 5th binhai road
wenzhou, zhejiang 32502 5
CH   325025
Manufacturer Contact
mohanapriya kannupaiyan
311 s. wacker dr
suite 4100
chicago, IL 60606
MDR Report Key22035458
Report Number3014312726-2025-00075
Device Sequence Number20996604
Product Code GAA
UDI-Device Identifier10884521544826
UDI-Public(01)10884521544826(10)02312068
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11811022
Device Lot Number02312068
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-