MAUDE Adverse Event Report: SOL-MILLENNIUM MEDICAL INC., CHICAGO, IL, USA. SOL-MILLENNIUM; SOL-M BLUNT FILL NEEDLE 18G*1 1/2"

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOL-MILLENNIUM MEDICAL INC., CHICAGO, IL, USA. SOL-MILLENNIUM; SOL-M BLUNT FILL NEEDLE 18G*1 1/2" Back to Search Results
Model Number 11022
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There is no evidence of quality issues associated with this complaint based on the return sample and archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 04/09/2025.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
 
Event or Problem Description
Customer reported that blunt needle leading to coring of rubber top vials as pictured in the attachment.The rubber debris and could be drawn up into a syringe and administered to a patient.We have tested a number of sol-m needles on different rubber top vials and they have core the rubber each time.
 
Additional Manufacturer Narrative
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.This supplemental mdr is being submitted to correct all manufacturer contact office information and date provided in the initial mdr [3014312726-2025-00095].The previously reported was incorrect.The correct manufacture contact information is below; (b)(4), qa engineer, sol-millennium medical inc, 311 s.Wacker dr., suite 4100, chicago, il 60606, united states (b)(4).The correct date of event is (b)(6) 2025.The complaint has been confirmed based on the customer provided picture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL-MILLENNIUM
Common Device Name
SOL-M BLUNT FILL NEEDLE 18G*1 1/2"
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC., CHICAGO, IL, USA.
311 s wacker dr, suite 4100
chicago IL 60606
Manufacturer (Section G)
ZHEJIANG KINDLY MEDICAL DEVICES CO., LTD
no.758 5th binhai road
wenzhou, zhejiang 32502 5
CH   325025
Manufacturer Contact
mohanapriya kannupaiyan
311 s. wacker dr
suite 4100
chicago, IL 60606
MDR Report Key22035487
Report Number3014312726-2025-00095
Device Sequence Number8655780
Product Code GAA
UDI-Device Identifier00818392015294
UDI-Public(01)00818392015294(10)02404030(10)02410028(10)05403025
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number11022
Device Lot Number02404030, 02410028, 05403025
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/25/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date08/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-