| Model Number |
11022 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There is no evidence of quality issues associated with this complaint based on the return sample and archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 04/09/2025.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
|
| |
|
Event or Problem Description
|
|
Customer reported that blunt needle leading to coring of rubber top vials as pictured in the attachment.The rubber debris and could be drawn up into a syringe and administered to a patient.We have tested a number of sol-m needles on different rubber top vials and they have core the rubber each time.
|
| |
|
Additional Manufacturer Narrative
|
|
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.This supplemental mdr is being submitted to correct all manufacturer contact office information and date provided in the initial mdr [3014312726-2025-00095].The previously reported was incorrect.The correct manufacture contact information is below; (b)(4), qa engineer, sol-millennium medical inc, 311 s.Wacker dr., suite 4100, chicago, il 60606, united states (b)(4).The correct date of event is (b)(6) 2025.The complaint has been confirmed based on the customer provided picture.
|
| |
|
Search Alerts/Recalls
|