Model Number BN1815
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There is no evidence of quality issues associated with this complaint based on the return sample and archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 06/03/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
 
Event or Problem Description
Customer reported: the needles took out fragments of the silicone in the vial when handled.There were fragments of silicone that remained floating in the medication and that could be aspirated during the loading of the dose by nurses and reach the patient.The occurrence of this event was during the immediate use of the product and there is no patient involvement.The sample is available for evaluation.As per the report there was no other pertinent product in use during the incident, the incident was not reported to any regulatory agency or competent body, and there was no reported harm or medical intervention required.
 
Additional Manufacturer Narrative
(b)(4) was initiated at the manufacturing site to track the investigation and application of corrective actions associated with the event.Corrective actions include optimization of sandblasting process and equipment, increase inspection sampling frequency, and update engineering drawings to clarify sandblasting requirements.Corrective actions were verified by inspection of 3 batchs produced after improvement implementation.
 
Additional Manufacturer Narrative
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.This supplemental mdr is being submitted to correct report type - adverse event and product problem provided in the initial mdr [3014312726-2024-00248].The previously reported type was incorrect.The correct report type is product problem only.(b)(4) was initiated at the manufacturing site to track the investigation and application of corrective actions associated with the event.Corrective actions include optimization of sandblasting process and equipment, increase inspection sampling frequency, and update engineering drawings to clarify sandblasting requirements.Corrective actions were verified by inspection of 3 batchs produced after improvement implementation.The first lot number produced after scar implementation is 02409011.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL-MILLENNIUM
Common Device Name
SMITHS MEDICAL BLUNT FILL NEEDLE 18G*1 1/2"
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC.
311 south wacker drive
suite 4100
chicago IL 60606
Manufacturer (Section G)
ZHEJIANG KINDLY MEDICAL DEVICES CO., LTD.
no. 758, 5th binhai road binha
longwan district
wenzhou, zhejiang province 32502 5
CH   325025
Manufacturer Contact
jarryd
311 south wacker dr.
suite 4100
chicago 60606
MDR Report Key22035518
Report Number3014312726-2024-00248
Device Sequence Number18306675
Product Code GAA
UDI-Device Identifier00304040017296
UDI-Public00304040017296
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
NA
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Other
Remedial Action Modification/Adjustment
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/14/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBN1815
Device Lot Number0221210
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/27/2024
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date05/28/2025
08/14/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-