| Model Number |
110022 |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
12/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There is no evidence of quality issues associated with this complaint based on the archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 12/18/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
|
| |
|
Event or Problem Description
|
|
Customer reported that they suffering a needle stick injury.
|
| |
|
Additional Manufacturer Narrative
|
|
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.Scar (b)(4) was initiated at the manufacturing site to track the investigation and application of corrective actions associated with the event.Corrective actions include optimization of sandblasting process and equipment, increase inspection sampling frequency, and update engineering drawings to clarify sandblasting requirements.Corrective actions were verified by inspection of 3 batchs produced after improvement implementation.This supplemental mdr is being submitted to correct report type - adverse event and product problem provided in the initial mdr [3014312726-2024-00378].The previously reported was incorrect.The correct report type is product problem only.
|
| |
|
Search Alerts/Recalls
|