SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-CARE SAFETY NEEDLE 25G*1 1/2" Back to Search Results
Model Number SN2515
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
There is no evidence of quality issues associated with this complaint based on the return sample and archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 08/22/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
 
Event or Problem Description
Customer reported: ahs mdip reference number (id): (b)(4).Date of incident (yyyy-mm-dd): (b)(6) 2024 incident details: item leaks at the hub when injected.Level of harm: no apparent harm - reached patient/person, inconvenient.
 
Additional Manufacturer Narrative
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL-MILLENNIUM
Common Device Name
SOL-CARE SAFETY NEEDLE 25G*1 1/2"
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC
311 s. wacker dr
suite 4100
chicago IL 60606
Manufacturer (Section G)
PROMISEMED HANGZHOU MEDITECH CO.,LTD
no.1388 cangxing street,
cangqian community
hangzhou city311121,
CH  
Manufacturer Contact
jarryd franklin
311 s. wacker dr
suite 4100
chicago, IL 60606
MDR Report Key22035607
Report Number3014312726-2024-00298
Device Sequence Number8778781
Product Code FMI
UDI-Device Identifier00818392010503
UDI-Public(01)00818392010503
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K123684
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Model NumberSN2515
Device Lot Number07105012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/21/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-