• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-CARE SAFETY WINGED INFUSION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-CARE SAFETY WINGED INFUSION SET Back to Search Results
Model Number 233412SI
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/18/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report was initially submitted on 09/05/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
 
Event or Problem Description
Customer reported: new butterfly needles poor quality, difficult to slide up safety needle.Manufacturer instructions were followed but the manufacturer of the blue winged butterfly needles has changed and staff has since found the safety cap more flimsy and difficult to manoeuvre.
 
Additional Manufacturer Narrative
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.Our evaluation of the risk, which follows guidance from iso 14971, indicates that the overall risk for the reported failure is low.This evaluation is based on our track and trend analysis and risk management files which are evaluated and updated constantly based on post market surveillance.Due to unavailability of batch number, the investigation couldn't able to conclude.As part of our continued commitment to our customers, sol millennium will continue to monitor the complaints trend for the reported failure mode and take any corrective action based on our procedures, if a significant pattern emerges.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL-MILLENNIUM
Common Device Name
SOL-CARE SAFETY WINGED INFUSION SET
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC
311 s. wacker dr
suite 4100
chicago IL 60606
Manufacturer Contact
jarryd franklin
311 s. wacker dr
suite 4100
chicago, IL 60606
MDR Report Key22035615
Report Number3014312726-2024-00302
Device Sequence Number8788338
Product Code JKA
UDI-Device Identifier00818392018295
UDI-Public(01)00818392018295
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K163073
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number233412SI
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/14/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date08/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-