| Model Number |
110022 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There is no evidence of quality issues associated with this complaint based on the return sample and archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 07/26/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
|
| |
|
Event or Problem Description
|
|
Customer reported: the customer advised that they have had numerous issues with the blunt needles getting stuck in the septum of medication vials, and then spraying staff when they are finally able to pull the needles out.The customer feels that the needles are a hazard to their staff.
|
| |
|
Additional Manufacturer Narrative
|
|
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.
|
| |
|
Search Alerts/Recalls
|