| Model Number |
100081IM |
| Medical Device Problem Codes |
Decoupling (1145); Suction Failure (4039)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report was initially submitted on 10/15/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
|
| |
|
Event or Problem Description
|
|
Customer reported: it is not possible to withdraw insulin.Almost nothing gets into the syringe itself, and suddenly the needle comes loose and ends up inside the syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.Our evaluation of the risk, which follows guidance from iso 14971, indicates that the overall risk for the reported failure is low.This evaluation is based on our track and trend analysis and risk management files which are evaluated and updated constantly based on post market surveillance.Due to unavailability of batch number, the investigation couldn't able to conclude.As part of our continued commitment to our customers, sol millennium will continue to monitor the complaints trend for the reported failure mode and take any corrective action based on our procedures, if a significant pattern emerges.
|
| |
|
Search Alerts/Recalls
|