MAUDE Adverse Event Report: SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-CARE 10ML LUER LOCK SAFETY SYRINGE W/EXCH NEEDLE 21G*

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-CARE 10ML LUER LOCK SAFETY SYRINGE W/EXCH NEEDLE 21G* Back to Search Results
Model Number TW160072IMLD
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.
 
Event or Problem Description
Customer reported upon opening the packaging, it was found that the needle had already retracted and was unusable.This issue has occurred multiple times and requires enhanced quality control.
 
Additional Manufacturer Narrative
There is no evidence of quality issues associated with this complaint based on the archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 01/30/2025.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
 
Additional Manufacturer Narrative
No samples were returned by the customer for analysis.In addition, investigation on retained samples did not confirm the defect.Regardless, scar (b)(4) was initiated at the manufacturing site to track the investigation and application of corrective actions associated with the event.Planned corrective action includes the implementation of a new "assemble condition inspection" matching to verify proper assembly of components.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL-MILLENNIUM
Common Device Name
SOL-CARE 10ML LUER LOCK SAFETY SYRINGE W/EXCH NEEDLE 21G*
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC
311 s. wacker dr
suite 4100
chicago IL 60606
Manufacturer (Section G)
ZHEJIANG LONGDE PHARMACEUTICAL CO., LTD.
no. 501, shunfeng rd
linping district, 311106
hangzhou, zhejiang,,
CH  
Manufacturer Contact
jarryd franklin
311 s. wacker dr
suite 4100
chicago 
MDR Report Key22035646
Report Number3014312726-2025-00014
Device Sequence Number8788352
Product Code MEG
UDI-Device Identifier00818392015041
UDI-Public(01)00818392015041
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K092430
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Remedial Action Modification/Adjustment
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTW160072IMLD
Device Lot Number04407049
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2025
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date05/28/2025
07/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-