Customer reported: please note that there is a product complaint from ministry of health, oman, regarding the product sol-m oral syringe.The product is oral syringe (lot number 19051447) which we supplied earlier has hair inside the pack.This is a serious issue and became a big discussion at moh.
|
|
The picture and sample received from the customer for evaluation of the impacted sol-m 10ml oral dispensing syringe amber w/tip cap, ref 20010oa, lot 19051447 confirmed the complained issue, residual/foreign material-blister (hair).However, no anomalies were found in all archived samples of the impacted sol-m 10ml oral dispensing syringe amber w/tip cap, ref 20010oa, lot 19051447 checked.Based on manufacturing partner feedback, the reported problem could have been caused by the carelessness of the operator who did not carefully wear protective clothing and hats as required, and by the inspection operator who did not detect the defective part in time, causing it to be released onto the market.The manufacturer partner has personnel to inspect products at every production stage (from the injection moulding, assembly, packaging, complete inspection and other processes).If defective products are detected, they are removed in time.For this reason, the complained issue is considered as an isolated case.As corrective action, the manufacturer partner changed the headwear in june 2024 to a new one with a better protective effect.The sol-millennium quality team checked the use of the new headwear in the production line by tracking the batches 01407018, 01407020 and 01407022 without encountering the complained issue.The new headwear has been verified to be effective without any new risk.Furthermore, the manufacturer partner retrained the related operators requiring all personnel to properly wear protective clothing and hats in designated areas before entering the production area; at the same time, cleaning staff should regularly clean the equipment and its work surface.The manufacturer partner retrained the full inspection staff and required each product to be checked one by one.If defective products with hair in the packaging are found, they should be removed in time to avoid entering the market.Sol-millennium will continue to monitor this kind of issue and in case the number of complaints increases, further evaluation and corrective actions will be taken.This report was initially submitted on 09/06/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
|