• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-M 60ML LUER LOCK SYRINGE W/O NEEDLE (BULK, STERILE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-M 60ML LUER LOCK SYRINGE W/O NEEDLE (BULK, STERILE) Back to Search Results
Model Number P180060BS
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No anomalies were found on archived and received samples of sol-m 60ml luer lock syringe w/o needle (bulk, sterile) ref (b)(4) lot 04402070 checked.Based on manufacturing partner feedback, the reported problem could have been caused by the carelessness of the operator who did not carefully wear protective clothing and hats as required, and by the inspection operator who did not detect the defective part in time, causing it to be released onto the market.The manufacturer partner has personnel to inspect products at every production stage (from the injection molding, assembly, packaging, complete inspection and other processes).If defective products are detected, they are removed in time.For this reason, the complained issue is considered as an isolated case.Manufacturer partner retrained the related operators requiring all personnel to properly wear protective clothing and hats in designated areas before entering the production area; at the same time, cleaning staff should regularly clean the equipment and its work surface.The manufacturer partner retrained the full inspection staff and required each product to be checked one by one.If defective products with hair in the packaging are found, they should be removed in time to avoid entering the market.Sol-millennium will continue to monitor this kind of issue and in case the number of complaints increases, further evaluation and corrective actions will be taken.This report was initially submitted on 10/04/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
 
Event or Problem Description
Customer reported: i would like to inform you that during our process we have found hair inside the syringe.
 
Additional Manufacturer Narrative
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL-MILLENNIUM
Common Device Name
SOL-M 60ML LUER LOCK SYRINGE W/O NEEDLE (BULK, STERILE)
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC
311 s. wacker dr.,
suite 4100
chicago IL 60606
Manufacturer (Section G)
ZHEJIANG LONGDE PHARMACEUTICAL CO., LTD.
no.510 shunfeng road
linping district hangzhou
zhejiang, 31110 6
CH   311106
Manufacturer Contact
jarryd franklin
311 s. wacker dr.,
suite 4100
chicago, IL 60606
MDR Report Key22035665
Report Number3014312726-2024-00321
Device Sequence Number8788365
Product Code FMF
UDI-Device Identifier00810018095239
UDI-Public(01)00810018095239
Combination Product (Y/N)N
Initial Reporter CountrySW
PMA/510(K) Number
K101359
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 08/12/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberP180060BS
Device Lot Number04402070
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2024
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date08/12/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-