MAUDE Adverse Event Report: SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-M BLUNT FILL NEEDLE 18G*1 1/2"

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOL-MILLENNIUM MEDICAL INC SOL-MILLENNIUM; SOL-M BLUNT FILL NEEDLE 18G*1 1/2" Back to Search Results
Model Number 110022
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
Customer complaint number: (b)(4).Customer reported that the rubber particles inside medicine vials and syringes.
 
Additional Manufacturer Narrative
There is no evidence of quality issues associated with this complaint based on the archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 01/30/2025.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
 
Additional Manufacturer Narrative
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.Scar 20241001 was initiated at the manufacturing site to track the investigation and application of corrective actions associated with the event.Corrective actions include optimization of sandblasting process and equipment, increase inspection sampling frequency, and update engineering drawings to clarify sandblasting requirements.Corrective actions were verified by inspection of 3 batchs produced after improvement implementation.This supplemental mdr is being submitted to correct the manufacturing site information provided in the initial mdr [3014312726-2025-00015].The previously reported site was incorrect.The correct manufacturing site is: zhejiang kindly medical devices co., ltd.No.758, 5th binhai road binhai industrial park longwan, wenzhou, zhejiang province.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOL-MILLENNIUM
Common Device Name
SOL-M BLUNT FILL NEEDLE 18G*1 1/2"
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC
311 s. wacker dr
suite 4100
chicago IL 60606
Manufacturer Contact
jarryd franklin
no. 758, 5th binhai road
binhai industrial park
longwan ,wenzhou,, zhejiang province 
CH  
MDR Report Key22035693
Report Number3014312726-2025-00015
Device Sequence Number8733631
Product Code GAA
UDI-Device Identifier00818392015294
UDI-Public(01)00818392015294
Combination Product (Y/N)N
Initial Reporter CountryPO
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number110022
Device Lot Number02401091
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/24/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date07/17/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/12/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-