| Model Number |
SN2510 |
| Medical Device Problem Codes |
Defective Component (2292); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/05/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Customer reported: needles have several issues.Needles are flimsy and bend easily.Safety cap doesnt snap in place cleanly and sometimes doesnt offer protection and needle hub doesnt hold to syringe well.
|
| |
|
Additional Manufacturer Narrative
|
|
There is no evidence of quality issues associated with this complaint based on the archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 08/08/2024.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
|
| |
|
Additional Manufacturer Narrative
|
|
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.
|
| |
|