| Model Number |
SGN2515 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
02/18/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
There is no evidence of quality issues associated with this complaint based on the return sample and archive sample analysis.Based on the determined risk priority number, the residual risk related to the complaint is considered acceptable.Sol millennium continues to monitor this kind of issue.Additional assessment will be taken if a significant pattern emerges.Unconfirmed.This report was initially submitted on 02/27/2025.Due to issues with displaying the initial submission, this submission contains both initial and supplemental information.
|
| |
|
Event or Problem Description
|
|
Customer reported that the needle broke off and was stuck in the patient.This is the second time this has happened to him.It looked to have broken off at the top where the needle meets the plastic safety.
|
| |
|
Additional Manufacturer Narrative
|
|
This report when initially submitted encountered an internal issue potentially related to database connectivity.This issue was identified while implementing capa-57 which was opened following an fda inspection.
|
| |
|
Search Alerts/Recalls
|