It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in the mildly tortuous and severely calcified right common iliac artery via ipsilateral approach.During the procedure, after the placement of the 6f non-bsc introducer sheath, intravascular ultrasound (ivus) was conducted following successful lesion crossing using a non-bsc guidewire.Pre-inflation was performed with an 8.0mmx60mmx135cm (4f) sterling balloon, and two epic self-expanding stents, were deployed.However, during post inflation using the same sterling balloon, the balloon ruptured upon fifth inflation at 12 atmospheres for 30-60 seconds.The device was removed without any problem, and post dilation was done with another of the same device.The procedure was completed with an ivus.There were no patient complications as a result of this event.
|