ABBOTT MEDICAL AMPLATZER MULTI-FENESTRATED SEPTAL OCCLUDER - CRIBRIFORM; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-MF-035
Medical Device Problem Code Difficult to Fold, Unfold or Collapse (1254)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It is unknown if the device is returning for analysis.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2025, a 35mm amplatzer cribriform occluder was selected for implant using a 10f amplatzer trevisio delivery sheath.During implant, when the occluder was deployed, the left disc deformed into a tulip shape.The occluder was reloaded, however the deformation persisted.There were no interactions with cardiac structures during deployment and no angulation/kink observed with the delivery system.The occluder was never released from the delivery cable while inside the patient.It was exchanged for a new 35mm amplatzer cribriform occluder, which was successfully implanted.The patient was reported to be in stable condition.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all defined manufacturing specifications.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER MULTI-FENESTRATED SEPTAL OCCLUDER - CRIBRIFORM
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key22038222
Report Number2135147-2025-02595
Device Sequence Number18187106
Product Code MLV
UDI-Device Identifier00811806010311
UDI-Public(01)00811806010311(17)290831(10)10534360
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/04/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9-ASD-MF-035
Device Catalogue Number9-ASD-MF-035
Device Lot Number10534360
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2025
Supplement Date Received by Manufacturer05/29/2025
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date06/04/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/25/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-