Catalog Number 333877
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 02/27/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: incident_description lors de l'anesthésie rachidienne, il y a eu contact osseux avec l'aiguille et lors du retrait pour se repositionner, découverte que l'aiguille est brisée en deux.Le bout de l'aiguille spinale a dû être retiré chirurgicalement.Translation: during the spinal anesthesia, there was bone contact with the needle and upon withdrawal to reposition, it was discovered that the needle was broken in two.The tip of the spinal needle had to be surgically removed.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Five retains from the reported lot number were tested and passed per specification.No sample was provided for evaluation.Based on the data from the investigation, the root cause of the reported incident was unable to be determined.The reported defect was unable to be confirmed.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENCAN®
Common Device Name
NEEDLE, CONDUCTION, ANESTHETIC (W/WO INTRODUCER)
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key22039101
Report Number2523676-2025-00521
Device Sequence Number9096171
Product Code BSP
UDI-Device Identifier04046963069568
UDI-Public(01)04046963069568(11)240529(17)290430(10)00VL945972
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K932569
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number333877
Device Lot Number00VL945972
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/16/2025
Supplement Report FDA Received Date07/09/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-