| Catalog Number |
MP1000 CHINA |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: no samples were received for investigation of (b)(4), in which the customer has stated: ¿during the use of this product, the piston joint failed to rebound ¿.The product in use at the time is reported to be a mp1000 china from lot 24015042.No further information was available to assist the investigation in this instance.The details of this feedback were forwarded to the manufacturing site for investigation, where a review of the production records for lot 24015042 did not identify any in-process testing failures or quality deviations which may have caused or contributed to a report of this nature.The root cause of the customer¿s experience could not be determined in this instance as no sample was received for investigation.Without a sample to examine it is not possible to determine whether a manufacturing defect could have caused or contributed to a report of this nature.In this instance, without the complaint sample to examine it has not been possible to conclusively link this feedback to a specific failure mode; however, a review of the customer feedback database indicates that this is a rare occurrence with a small number of similar reports against the mp1000 china product in the past 12 months.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd maxplus positive pressure connector was damaged.The following information was provided by the initial reporter, translated from chinese to english: during the use of this product, the piston joint failed to rebound.
|
| |
|
Search Alerts/Recalls
|