|
As reported, not all of the filter was in the "garage" and the tip of the 5mm basket, medium support, angioguard rx emboli capture guidewire system was already pre curved.It appeared to prep fine, but when it was taken out of its loop, not all of the filter was in the garage, and the tip of the wire was already curved.The physician attempted to re-prep the filter.It would go into the garage, but the filter would not completely stay there unless resistance was applied to the wire.A 4mm angioguard filter was successfully prepped and used to complete the case.There were no reports of patient injury.There were no damages found on the device packaging prior to opening.The device was stored and prepped in accordance with the instructions for use (ifu).The device was returned for evaluation.Addendum: per pe preliminary findings, the distal tip is presenting a waved (out of shape) and unraveled condition.
|
|
Complaint conclusion: as reported, not all the filter was in the "garage" and the tip of the 5mm basket, medium support, angioguard rx emboli capture guidewire system was already pre curved.It appeared to prep fine, but when it was taken out of its loop, not all the filter was in the garage, and the tip of the wire was already curved.The physician attempted to re-prep the filter.It would go into the garage, but the filter would not completely stay there unless resistance was applied to the wire.A 4mm angioguard filter was successfully prepped and used to complete the case.There were no reports of patient injury.There was no damage found on the device packaging prior to opening.The device was stored and prepped in accordance with the instructions for use (ifu).The device was returned for evaluation.A non-sterile unit of a ¿5mm basket, medium support, angioguard rx for us carotid¿ was received in a clear plastic bag.Upon unpacking, the returned components included the deployment sheath, the capture sheath, and the ecgw system; the remaining components were not returned.The ecgw system was received inserted into the deployment sheath.A kink was observed on the guidewire approximately 23 cm from the proximal end, and the distal tip exhibited an out-of-shape and unravelling condition.The filter basket area was inspected using a vision system, which revealed no anomalies or damage to the filter struts or membrane walls.Functional analysis was performed using lab samples of the filter introducer, torquing device, and coil dispenser.The ecgw system was removed from the delivery sheath, inserted into the filter introducer, and reinserted into the delivery sheath.The loaded sheath was then introduced into the dispenser coil, with the filter secured to it.A torquing device was attached to the guidewire¿s proximal end, and the guidewire was pulled until the filter basket docked fully into the tip of the deployment sheath.Visual inspection confirmed the kinked guidewire and the distal tip¿s unravelling, out-of-shape condition.The reported ¿delivery system-loading (docking) difficulty-into delivery sheath¿ was not observed during the product analysis.The filter basket was fully docked into the delivery sheath during functional analysis.However, the reported ¿distal tip (ecgw)-out of shape¿ was observed and caused by the observed malfunction ¿distal tip (ecgw)-unraveled/stretched.¿ the unraveled condition resulted in an out of shape distal tip.The exact cause of these events cannot be determined but may have been caused by handling factors encountered while preparing the device for use.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent of making the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿caution: guidewires are delicate instruments and should be handled carefully.Prior to use, and when possible, during the procedure, inspect the guidewire carefully for coil separation, bends, kinks, or damage of the filter basket assembly.Do not use defective equipment.Fill a 5 ml luer lock syringe with sterile saline and purge all air.Attach syringe to luer lock hub on the end of the filter introducer.Inject 5 ml of saline to purge all air from the deployment sheath and filter basket (ensure distal tip of the deployment sheath is inside the filter basket introducer tip prior to purging the system).You should see saline dripping from the proximal end of the deployment sheath (inside the coil dispenser).Disconnect syringe.Remove the two proximal (closest to the torque device) anti-migration clips holding the guidewire in the dispenser coil.Ensure the torque device is locked onto the guidewire.Gripping the torque device in one hand and the coil dispenser in the other, pull on the wire until the basket is completely docked into the tip of the deployment sheath.When completely docked, approximately half the filter basket will still be visible out of the end of the deployment sheath.After the deployment sheath is completely docked, remove the remaining distal anti-migration clip (closest to the filter introducer) and continue to pull back on the torque device to disengage the docked deployment sheath/guidewire assembly from the filter introducer.Loosen the torque device.While holding the torque device in one hand, gradually pull back on the guidewire with the other hand.Continue to pull the guidewire through the torque device until the proximal end of the deployment sheath is visible at the proximal end of the torque device.Tighten the torque device onto the deployment sheath to secure the deployment sheath to the guidewire.¿ based on the available information, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventative or corrective actions will be taken at this time.
|