RESPIRONICS, INC. BIPAP A40 PRO; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
| Model Number |
CAX3100S12 |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/19/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a ventilator inoperative condition occurred.There was no harm or injury reported.The ventilator was returned to the manufacturer's product investigation laboratory (pil) for evaluation and the customer¿s complaint was confirmed pil performed testing in accordance with er 2251830 part 002 ver 01 and the results meet the criteria that no further investigation is required.E-20 error was generated due to a defective eeprom.The device has been scrapped.Prior to identifying the issue described in fsn 2023-cc-src-039, complaints associated with the issue did not meet requirements for vigilance reporting as a reportable incident.As part of the hhe process, a retrospective review of complaints associated with the issue described in fsn 2023-cc-src-039 was necessary to reassess reportability.Upon this further assessment, these complaints are being reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The bipap a40 pro (cax3100s12) is substantially similar to the bipap a40 and will be reported in the united states under bipap a40, 510k number: k121623.
|
| |
|