• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I T3 PRO NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25MM (D) X 10MM (L); DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I T3 PRO NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25MM (D) X 10MM (L); DENTAL IMPLANT Back to Search Results
Catalog Number T3ST3210
Medical Device Problem Codes Failure to Osseointegrate (1863); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Sinus Perforation (2277); Osteolysis (2377)
Date of Event 03/18/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a non integration due to a sinus perforation along with infection and bone loss at tooth site 26.Implant was located into sinus and later removed, after 4 months.Infection detected around implant before loading it.Implant was removed.Symptoms reported on the patient: inflammation and pain.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).H6: event problem codes ¿ patient code: 1994 pain, 1932 inflammation.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H10: additional narrative.Zimvie received one (1) t3st3210, (t3 pro non-platform switched tapered implant 3.25mm (d) x 10mm (l)) for evaluation.Visual evaluation was performed, the implant had signs of use with debris on its external threads.There was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Device history record (dhr) review was completed for the subject lot number 2120004888.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120004888 for similar events and no other complaint was identified.Review completed utilizing keywords: dental: functional: non-integration: perforated sinus based on the investigation and risk management file review, the most likely root cause determined from the investigation was customer error and patient factors.For infection and bone loss.A summary investigation has been completed for bone loss and infection events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for bone loss and infection have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of manufacturing or design defects that might lead to bone loss and escaping the available detections has been assessed and found remote and almost nonexistent.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable with all the available information provided.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
T3 PRO NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25MM (D) X 10MM (L)
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key22042419
Report Number0001038806-2025-01110
Device Sequence Number12981730
Product Code DZE
UDI-Device Identifier00844868046868
UDI-Public(01)00844868046868(17)281017(10)2120004888
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K213672
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 06/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue NumberT3ST3210
Device Lot Number2120004888
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/30/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/19/2025
Supplement Report FDA Received Date06/20/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/19/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-