Additional information - this mdr is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary: the information in this complaint (b)(4) has been evaluated.The batch 6008621 in question was manufactured at the reynosa site.Investigation process of the complaint was carried out in accordance with work instruction (wi) guidance for visual testing for complaints area version 3 and wi guidance for functional testing for complaints area version 2 for the code occlusion in the tubing and/or tubing connectors (e.G.Occlusion alarm from the infusion pump may have sounded).Complaint investigation.Photo/sample was not provided.In order to test the product, the reference samples from the lot have been requested.Reference samples test results: on the visual inspection according to wi version 3, was performed and 10/10 samples passed the test.Functional test air flow according to wi version 2 on reference samples, 10/10 samples passed the test.A batch record review was conducted resulting in the following: packaging lot: the lot 6008621 was packaging according to the wi version 19, in the machine multivac 14, on 06/aug/2024, with a total of (b)(4) units.Glue-tubing lot: the lot 4g06294 was manufactured according to the wi version 15, in the machine lc01, on 01/aug/2024, with a total of (b)(4) units.Review of the device history record (dhr) showed that all relevant tests required during the related processes had been fulfilled and met the requirements.No deviation were identified; no maintenance events were recorded.Trending: a query was run in database on 24-jun-2025 against malfunction code occlusion in the tubing and/or tubing connectors (e.G.Occlusion alarm from the infusion pump may have sounded) and lot 6009739 and one other complaint has been registered in database for the same lot and malfunction code.Conclusion summary of the related event: as a result of the following: on the investigation on reference samples, no issue were found related to occlusion in tubing, harm no reportable, no defect on tests, no non-conformance (nc) raised during production, no more complaint was received on the lot in question and malfunction, no further action are required, therefore, this issue will be monitored through the post market surveillance activities.
|