| Catalog Number |
8065753075 |
| Medical Device Problem Code |
Improper Flow or Infusion (2954)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A customer reported that an ophthalmic system exhibited an issue with the irrigation fluid that was directly going to drain bag.Procedure details and patient impact were not reported.
|
| |
|
Event or Problem Description
|
|
Additional information received stating that the issue was observed during the cataract surgery in the quad mode.The surgery was completed with the another system with no impact on patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information is provided in sections b.3, b.5, d.9, h.3, h.6 and h.11.The company representative was able to replicate the reported event.The nonconforming fluidics was replaced to address the issue.The system was tested and found to meet product specifications.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation with the same event code the root cause of the reported event is attributed to nonconforming fluidics.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
|
| |
|