Additional information: h4, h6 correction: it was determined after receiving further clarification from the customer and the log files from the device, this event for this product was not a reportable event per cfr 21, part 803.This supplemental is being filed to identify an mdr was not required for this event.As per the information received this device is concomitant.It did not cause or contribute to this event.The reported event could not be confirmed, because neither the products nor any further information could be provided.A review of all relevant manufacturing and inspection records for the reported screws and plate found no issues, and the lots were properly manufactured and accepted into final stock.According to the risk file, such events may result from user-related factors or surgical conditions, including incorrect implant selection, improper assembly or insertion, and patient-specific limitations.Based on the investigation, there is no indication for an incorrectly working product or any design, material, or manufacturing-related issue.Therefore, no corrective and/or preventive actions are deemed necessary at this time.
|