BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. BD INSYTE AUTOG BC WING; PERIPHERAL IV CATHETERS Back to Search Results
Catalog Number 382647
Medical Device Problem Codes Fluid/Blood Leak (1250); Retraction Problem (1536)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that bd insyte autog bc wing needle retraction slow resulting in blood leak.The following information was provided by the initial reporter: needle retraction sluggish or grinding pulling back very slowly allowing blood to leak from hub.
 
Additional Manufacturer Narrative
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
 
Additional Manufacturer Narrative
Since no samples displaying the condition reported were available for examination, we were unable to fully investigate this incident.We would be very interested in examining product that does not meet your expectations and our quality standards.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.Examination of the actual product involved may provide clarification as to the cause for the reported failure.A device history record review showed no non-conformances associated with this issue during the production of this batch.
 
Event or Problem Description
No new information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD INSYTE AUTOG BC WING
Common Device Name
PERIPHERAL IV CATHETERS
Manufacturer (Section D)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer (Section G)
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
9450 south state street
sandy UT 84070
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key22045897
Report Number1710034-2025-00857
Device Sequence Number8990437
Product Code FOZ
UDI-Device Identifier00382903826476
UDI-Public(01)00382903826476(11)240102(17)261130(10)3353170
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K201075
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number382647
Device Lot Number3353170
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2025
Supplement Date Received by Manufacturer09/09/2025
Initial Report FDA Received Date05/19/2025
Supplement Report FDA Received Date09/10/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-