| Model Number |
CF-Q165I |
| Medical Device Problem Code |
Microbial Contamination of Device (2303)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/12/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during reprocessing, the colonovideoscope tested positive for an unknown number of colony forming units (cfus) of stenotrophomonas maltophilia and tested positive for >60 (cfus) of aerobic micro-organisms at 30° c.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: no additional information received from customer.D8: additional information d9: additional information h2: additional information, device evaluation h3: additional information h4: additional information h6: additional information this report is being supplemented to provide additional information based on the legal manufacture's (lm) review of the customer cleaning disinfection and sterilization (cds) processes, the device evaluation and the lms final investigation.The lm reviewed the customer provided cds processes where information to judge deviation from the ifu was not identified.The device was evaluated where no abnormalities were found that could have led to the reported event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation and because the user culture results were not shared, the reported event could not be confirmed and a root cause could not be determined.Growth of microorganisms were reported through culture testing by the user after reprocessing.However, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No additional information received from customer.
|
| |
|
Search Alerts/Recalls
|