| Model Number |
CCW0T5 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Impairment (2138)
|
| Date of Event |
04/15/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced reason as blurry vision and clinical reason for explant being residual astigmatism more than measured.The iol was explanted and exchanged for an unknown advanced technology intraocular lens following the initial implant procedure.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction information has been provided in b.5., and h.11.Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury/malfunction.No further reports required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information has been requested and received stating that the iol was exchanged for the different lens model but half a diopter more power indicating the correct lens power was not implanted initially.There is no reported defect or fault with the iol.
|
| |
|
Search Alerts/Recalls
|