DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X30MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Catalog Number 130790030
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that patient underwent arthroplasty on (b)(6) 2025.The crown part of the screw got chipped off while tightening it.Procedure was completed successfully with no delay.There was no patient consequence.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the crown part of the screw got chipped off while tightening it.The product was not returned to j&j medtech orthopaedics; however, photos were provided for review.See attachment (b)(4).The photograph attached was reviewed, however does not represent the reported fracture, only the implant box was shown in the photographic evidence.Therefore, the investigation could not draw any conclusions about the reported event due to the insufficient evidence provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the photographs provided are not representative of the reported complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.H11 additional narrative: added: d10 (concomitant).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X30MM
Common Device Name
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key22046167
Report Number1818910-2025-08132
Device Sequence Number18396648
Product Code HSD
UDI-Device Identifier10603295027843
UDI-Public(01)10603295027843(17)20290630(10)5741494
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K192855
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number130790030
Device Lot Number5741494
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/27/2025
Supplement Date Received by Manufacturer05/20/2025
Initial Report FDA Received Date05/19/2025
Supplement Report FDA Received Date05/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DELTA CER HEAD 12/14 32MM +1
Patient SexUnknown
-
-