• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. VCL + CT BRD UD 8X18IN USP3-0 S/A SHB CR; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL + CT BRD UD 8X18IN USP3-0 S/A SHB CR; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCPB864D
Medical Device Problem Code Delivered as Unsterile Product (1421)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).H3 investigation summary: the product was returned to ethicon for evaluation.One open sample was received for analysis.The product code is vcpb864d.During the initial inspection of the sample, the foil packet received shows a hole that is located on the top foil.The hole appears to be caused outside in by an unknown object affecting the integrity of the sterility.Additionally, a photo was provided, and it showed a foil packet with a circle indicating the hole.Due to the damages found on the device, the complaint is confirmed.The condition of the package is indicative of handling and storage issues that occurred outside of ethicon control.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2025 and suture was used.Before use on the patient, it was reported that the primary package of the suture was found damaged prior to use.Changed another one to complete the surgery.There is no report on patient's injury.Enclosed please find defect photo.No additional information could be provided.There were no adverse consequences reported.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL + CT BRD UD 8X18IN USP3-0 S/A SHB CR
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22047807
Report Number2210968-2025-05544
Device Sequence Number18367546
Product Code GAM
UDI-Device Identifier10705031056664
UDI-Public(01)10705031056664(17)290831(10)103JXK
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 05/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCPB864D
Device Lot Number103JXK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/13/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2025
Initial Report FDA Received Date05/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/23/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-