| Catalog Number |
TPT1000 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
05/06/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that upon opening the glass ampule of lidocaine during thoracentesis and paracentesis use, the top of the ampule crushed causing glass shards and contaminated the sterile tray.It was further reported that the lidocaine vial was breaking into small fragments and caused a finger laceration on the user requiring intervention, user washed hands and applied a bandage.No other care was needed.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the lot number for the device was provided.The device history records are currently under review.The device is pending return.The investigation is currently underway.H10: b3: date of event: the date of awareness was entered in the date of event field as the actual date of event is unknown.H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: the physical sample was not received by our quality team for evaluation.Three photos were provided of the tray packaging, and one photo was provided of possible user storage rack and reviewed; but the quality team was unable to make any observations regarding reported failure based on photos provided.Therefore, the failure mode could not be verified, and the root cause could not be determined.A review of the device history record for part: tpt1000 lot: 0001604817 was performed.No recorded quality problems or rejections related to this incident were found.It was confirmed that procedural and functional requirements needed for its release were met.The supplier was notified of this reported failure to raise awareness.B3: date of event: the date of awareness was entered in the date of event field as the actual date of event is unknown.G3 (date received by manufacturer), g6 type or report - follow up# 1), h2 (correction and additional information), annex g (component code), annex b (type of investigation), annex c (investigation findings), annex d (investigation conclusions).Section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that upon opening the glass ampule of lidocaine during thoracentesis and paracentesis use, the top of the ampule crushed causing glass shards and contaminated the sterile tray.It was further reported that the lidocaine vial was breaking into small fragments and caused a finger laceration on the user requiring intervention, user washed hands and applied a bandage.No other care was needed.
|
| |
|