The customer stated that the qc was acceptable.The investigation confirmed the customer's elecsys total psa (total psa) results.The patient sample was tested with an additional psa antibody of another specificity.The result showed a significant increase in psa at up to 0.447 ng/ml, indicating the presence of a rare psa isoform that cannot be completely detected by the standard elecsys total psa (total psa) assay.Elecsys total psa (total psa) product labeling states, "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.It is known that in rare cases, psa isoforms do exist that may be measured differently by different psa tests.Findings of this kind have occasionally been reported for psa tests from various manufacturers.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings." elecsys free psa product labeling states, "in rare cases, interference due to extremely high titers of antibodies to analyte-specific antibodies, streptavidin or ruthenium can occur.These effects are minimized by suitable test design.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings." the investigation determined that an interferent and rare isoform had caused the event.The investigation did not identify a product problem.
|