The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that customer stated that the tubing on the meter bags was came out of the packaging kinked which was prevented the urine from draining.Customer recently explained that the drain tubing kickstand guide piece at the top of the bags was only meant for the tubing to go through on non urine meter drain bag foley trays.As they can see in the first two photos above, the bag on the right did not have a urine meter and the tubing was not positioned within the holder.They would be auditing that component and would let them know if they find additional bags without urine meters that did not had tubing in the holder as they mentioned.Additionally, they were tubing that was kinked because of the positioning during packaging.The plastic that was used had memory, and the second set of photos shows tubing that was kinked and was not drained well.When they pressed the tubing open, the urine began to drain.The care team was notified and an rl will be placed for that catheter.Ip will recommend a new drainage bag be placed.
|
|
The reported event was confirmed cause unknown.Visual evaluation of the returned two-photo sample noted one opened (without original packaging), used urine meter drainage bag.Visual inspection of the photo sample noted the outlet tubing was kinked at the top of the urine meter.No actions can be taken at this time since a root cause was not identified.The device history record review could not be performed without a lot number.The instructions for use were found adequate and state the following: 1.Occlude drainage tubing a minimum of 3 inches below the sampling port by kinking the tubing until urine is visible under the access site.2.Swab surface of bard ez-lok sampling port with antiseptic wipe.3.Using aseptic technique, position the syringe in the center of the sampling port.Press the syringe firmly and twist gently to access the sampling port.4.Slowly aspirate urine sample into syringe and remove syringe from sample port.5.Unkink tubing if necessary and transfer urine specimen into specimen cup or follow hospital protocol.Discard syringe according to hospital protocol.6.Follow established hospital protocol for specimen labeling and transport to lab.Sterile unless package is opened or damaged, except for any individually packaged components within the tray which are not labeled as sterile.These components are not terminally sterilized.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Single use only.Do not resterilize.For urological use only.Warning: on catheter, do not use ointments or lubricants having a petrolatum base.They will damage the catheter and may cause balloon to burst.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and applicable local, state and federal laws and regulations.Visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Please consult product label and insert for any indications, contraindications, hazards, warnings, cautions and directions for use.Proper techniques for urinary catheter insertion perform hand hygiene immediately before and after insertion insert urinary catheters using aseptic technique and sterile equipment use the smallest foley catheter possible, consistent with good drainage document the indications for catheter insertion, date and time of catheter insertion, individual who inserted catheter, and date and time of catheter removal in patient record proper techniques for urinary catheter maintenance secure the foley catheter, use the statlock foley stabilization device if provided maintain a closed drainage system by utilizing pre-connected, sealed catheter-tubing junctions maintain unobstructed urine flow and keep the catheter and collection tube free from kinking keep the collection bag below the level of the bladder or hips at all times empty the collection bag regularly (e.G., prior to transport) using a separate, clean collection container for each patient routine hygiene (e.G., cleansing of the meatal surface during daily bathing or showering) is appropriate leave foley catheter in place only as long as needed.Correction: h udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
It was reported that customer stated that the tubing on the meter bags came out of the packaging kinked which was prevented the urine from draining customer recently explained that the drain tubing kickstand guide piece at the top of the bags was only meant for the tubing to go through on non urine meter drain bag foley trays.As they can see in the first two photos above, the bag on the right did not have a urine meter and the tubing was not positioned within the holder.They would be auditing that component and would let them know if they find additional bags without urine meters that did not had tubing in the holder as they mentioned.Additionally, they were tubing that was kinked because of the positioning during packaging.The plastic that was used had memory, and the second set of photos shows tubing that was kinked and was not drained well.When they pressed the tubing open, the urine began to drain.The care team was notified, and an rl will be placed for that catheter.Ip will recommend a new drainage bag be placed.
|