INSPIRE MEDICAL SYSTEMS INSPIRE MODEL 4340 RESPIRATORY SENSING LEAD Back to Search Results
Model Number 4340
Medical Device Problem Code Positioning Problem (3009)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/05/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Patient presented to the physician with rib pain localized at the site of the placed sense lead.Physician brought the patient into the or, removed the original sense lead, placed a new sense lead in an alternate location, re-sutured, and closed the incision site.This resolved the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRE MODEL 4340 RESPIRATORY SENSING LEAD
Common Device Name
INSPIRE MODEL 4340 RESPIRATORY SENSING LEAD
Manufacturer (Section D)
INSPIRE MEDICAL SYSTEMS
5500 wayzatablvd.
suite 1600
golden valley MN 55416
Manufacturer (Section G)
INSPIRE MEDICAL SYSTEMS
5500 wayzata blvd.
suite 1600
golden valley MN 55416
Manufacturer Contact
vang yang
5500 wayzata blvd.
suite 1600
golden valley, MN 55416
7632057970
MDR Report Key22053002
Report Number3007666314-2025-00303
Device Sequence Number17481744
Product Code MNQ
UDI-Device Identifier10855728005632
UDI-Public10855728005632
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P130008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number4340
Device Catalogue Number900-009-002
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/05/2025
Initial Report FDA Received Date05/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-