| Model Number |
8404KXC |
| Medical Device Problem Codes |
Poor Quality Image (1408); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the light source was dim and failing.No patient or procedure involvement.No negative impact in state of health reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is not yet completed, investigation results are pending.This event is filed under internal complaint id: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturer's evaluation results: the device was sent to the manufacturer for inspection.During the manufacturer's technical inspection, the error description provided by the customer was confirmed.The device passed the quality check at the time of delivery the technical examination of the video laryngoscopes revealed several user-induced defects that led to a failure of the light output.Due to damage to the blade, and a leak between the plug and the cover, liquid can penetrate the interior of the housing where the electronics are located.This damages the circuit board due to a short circuit, which impairs the function responsible for light transmission and leads to a failure of the light output.The above conducted investigations did not reveal a root cause related to the labelling, the device or its history.All production related quality checks were passed, and no non-conformities were identified in the devices manufacturing records.The labelling was found to contain adequate instructions and warnings.This event is filed under internal complaint id (b)(4).
|
| |
|