• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RIVERPOINT MEDICAL LLC CHROMIC GUT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RIVERPOINT MEDICAL LLC CHROMIC GUT Back to Search Results
Model Number 558CG
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/08/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned.Three retained samples from the same lot were used to perform a knot pull tensile test.The retained samples used for the knot pull tensile test met requirements.Production records were reviewed.There were no nonconformities noted with the lot during production.All finished goods testing requirements were met prior to release.There was no evidence that the device failed to meet specifications, and the report could not be substantiated.A cause for the event cannot be established.This report and use of categorical definitions required by fda 3500a does not constitute an admission by riverpoint medical or its employees that riverpoint medical or its employees have caused or contributed to the event described in this report.Riverpoint medical has filed this information to comply with the medical device reporting regulation 21 cfr 803.If additional information is provided to riverpoint medical regarding this event, a supplementary 3500a form will be submitted as required by the fda.
 
Event or Problem Description
According to the reporter, "suture breaking easily and brittle.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHROMIC GUT
Common Device Name
CHROMIC GUT
Manufacturer (Section D)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
Manufacturer (Section G)
RIVERPOINT MEDICAL LLC
825 ne 25th ave
portland OR 97232
Manufacturer Contact
edwin anderson
825 ne 25th ave
portland, OR 97232
MDR Report Key22053445
Report Number3006981798-2025-00024
Device Sequence Number5228391
Product Code GAL
UDI-Device Identifier10790986001777
UDI-Public10790986001777
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K001299
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/20/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number558CG
Device Lot Number240131-27
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/08/2025
Initial Report FDA Received Date05/20/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-