| Catalog Number |
112062 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information: h6, h11.H11: the actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the photos showed the filter in the machine holder with a puddle of fluid visible below the filter.The specific defect that allowed the leak could not be identified in the photos.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the reported leakage was not determined as no further or actual sample testing could be conducted.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an external fluid leak was observed from an ultrafilter u9000 during an unspecified process step.The filter was used in an ak98 machine 123 times in 47 days.There was no patient involvement.No additional information is available.
|
| |
|