| Catalog Number |
03-2722-9 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
04/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
The nurse administrator from a user facility reported that a combi set blood leak occurred during a patient¿s hemodialysis (hd) treatment.The leak was caused by a detachment of the combi set line from the patient¿s dialysis access (not a fresenius product).It is unknown how much blood loss this resulted in, or if the patient was able to complete their treatment.There were no allegations made that the event resulted in any serious patient injury or the need for medical intervention.Multiple attempts were made to obtain additional information, and thus far no further details have been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: the sample was not returned to the manufacturer and the lot number was not provided.A manufacturing review was performed on the products shipped to the facility for the three (3) month time frame which immediately preceded the event occurrence date.This review included the lot numbers for all fresenius combi sets shipped to this account within the selected time frame.The entire set of lots have been sold and distributed.There were no non-conformances or abnormalities identified during the manufacturing process which could be associated with the reported event.An investigation of the device history records (dhr) was conducted and confirmed that the results of the in-progress and final quality control (qc) testing met all requirements.The product lots involved met all specifications for release.A review of the dhr did not reveal a probable cause for the customer complaint.As a physical evaluation could not be performed, a definitive conclusion regarding the reported incident could not be reached and a cause could not be confirmed.
|
| |
|
Event or Problem Description
|
|
The nurse administrator from a user facility reported that a combi set blood leak occurred during a patient¿s hemodialysis (hd) treatment.The leak was caused by a detachment of the combi set line from the patient¿s dialysis access (not a fresenius product).It is unknown how much blood loss this resulted in, or if the patient was able to complete their treatment.There were no allegations made that the event resulted in any serious patient injury or the need for medical intervention.Multiple attempts were made to obtain additional information, and thus far no further details have been provided.
|
| |
|