COVIDIEN SURGICAL TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIGTRSB45AMT
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/13/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant products: egiaustnd egiaustnd endogia ultra univ std stap - (lot#p4m0944); sigtrsb45amt sigtrsb45amt 45mm med thk reinforced rel - (lot#n3c0737y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a thoracoscopic wedge resection of the lung, after the nurse loaded the handle with the reload for the surgeon, the reload magazine fell off the handle.A second reload was loaded into the same handle.The surgeon placed the fully articulated reload on the lung tissue.The green button was pressed, and the instrument was fully fired, and the staples were placed in the tissue.Then, the black retraction button could not be retracted to open the instrument.The jaws locked and could not be removed from the tissue.The surgical procedure was converted into an open surgery, and the surgical time was extended by more than 60 minutes.The surgeon placed another stapler with a 45 mm purple loading unit distal to the blocked loading unit and then released the instrument.The blocked magazine was then removed from the site.The incision had to be widened considerably in the thoracic re gion.
 
Additional Manufacturer Narrative
Correction: b3 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: d3 (mfr name and address), d4 (udi#), d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the reload was fully fired.The knife bar was advanced to the distal end and the jaws were clamped.The proximal end of the reload adapter was broken.It was reported that the staples were placed in the tissue and the black retraction button could not be retracted.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur due to over flexure of the reload while attached to the instrument.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key22056039
Report Number1219930-2025-02218
Device Sequence Number4250989
Product Code GDW
UDI-Device Identifier10884521717602
UDI-Public(01)10884521717602(17)260228(10)N3C0737Y(20)02
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K173270
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Model NumberSIGTRSB45AMT
Device Catalogue NumberSIGTRSB45AMT
Device Lot NumberN3C0737Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2025
Supplement Date Received by Manufacturer07/22/2025
08/19/2025
Initial Report FDA Received Date05/20/2025
Supplement Report FDA Received Date08/08/2025
08/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-