| Model Number |
SIGTRSB45AMT |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/13/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: egiaustnd egiaustnd endogia ultra univ std stap - (lot#p4m0944); sigtrsb45amt sigtrsb45amt 45mm med thk reinforced rel - (lot#n3c0737y) medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a thoracoscopic wedge resection of the lung, after the nurse loaded the handle with the reload for the surgeon, the reload magazine fell off the handle.A second reload was loaded into the same handle.The surgeon placed the fully articulated reload on the lung tissue.The green button was pressed, and the instrument was fully fired, and the staples were placed in the tissue.Then, the black retraction button could not be retracted to open the instrument.The jaws locked and could not be removed from the tissue.The surgical procedure was converted into an open surgery, and the surgical time was extended by more than 60 minutes.The surgeon placed another stapler with a 45 mm purple loading unit distal to the blocked loading unit and then released the instrument.The blocked magazine was then removed from the site.The incision had to be widened considerably in the thoracic re gion.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b3 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d3 (mfr name and address), d4 (udi#), d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the reload was fully fired.The knife bar was advanced to the distal end and the jaws were clamped.The proximal end of the reload adapter was broken.It was reported that the staples were placed in the tissue and the black retraction button could not be retracted.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur due to over flexure of the reload while attached to the instrument.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|