SMITH & NEPHEW, INC. JOURNEY DCF AP FEM CUT BLK 4; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
74012414 |
| Medical Device Problem Codes |
Break (1069); Crack (1135)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: case-(b)(4).H3,h6:this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during tka, one (1) journey dcf ap fem cut blk 4 was broken.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported as a consequence of this issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.Device damage (cracks) may cause minor or temporary injury (e.G.Abrasion, laceration, etc.) requiring no or minor medical intervention.Event may result in a short procedural delay and/or require use of a backup device to continue procedure.This event is not likely to cause or contribute to a death or serious injury and is considered not reportable pursuant to applicable regulations.The associated device was returned and evaluated.A visual inspection of the returned device finds the device returned with one of the bars on the back cracked.The entire device is worn from use.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed part number over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions that could be related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|