ETHICON INC. PRLNE HS BLU 36IN 4-0 D/A RB-1; SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE Back to Search Results
Catalog Number HS6857H
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2025 and suture was used.It was reported that the problem of pulling off suture needle happened in surgery.Changed another three to continue the surgery, the same problem happened again.Changed another one to complete the surgery.There were no patient consequences reported.There will be 21 devices returned for analysis, including 4 actual product and 17 sterile products from same batch.The surgeon refused to use the remaining 17 sterile products from the same lot and they will be returned for analysis.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint #(b)(4).Date sent to the fda: 5/21/2025.H6 component code: g07002 - device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).H6: component code: g07002 - no device problem found.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H3 investigation summary: the product was returned to ethicon for evaluation.Three unused open samples were received for analysis.Product code hs6857h.During the visual inspection of the returned samples, the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformance's were identified.Expiration date: additional h3 investigation summary: the product was returned to ethicon for evaluation.One open box with seventeen unopened samples were received for analysis.As per the sampling plan, a visual inspection was performed on thirteen samples, the packets were opened.The swage and attachment area were noted as expected.The sutures were dispensed without problems and examined along the strand no anomalies were observed during the evaluation.The functional test was performed using instron equipment and the pull force result was above the minimum requirements.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory anal.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h6, d9, h3.H3 investigation summary: the product was returned to ethicon for evaluation.One winding former with a detached needle and one needle suture piece outside the overwrap packet were received for analysis.Product code hs6857h.Upon visual inspection of the detached needle, the swage and attachment area was noted to be as expected.The barrel hole was examined under magnification, and no suture remnant was noted.In addition, the needle suture piece was visually inspected, and the swage and attachment area was noted to be as expected.The suture was examined along of the strand and the end was noted to be cut probaby cause by a surgical instrument.In additon the other section of the suture was not returned for evaluation.A functional test was performed using instron equipment and the pull force was above the minimum requirements.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.No conclusion could be reached as to what caused the reported complaint.As part of the quality process, all devices are manufactured, inspected, and released to approved specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRLNE HS BLU 36IN 4-0 D/A RB-1
Common Device Name
SUTURE, NONABSORBABLE, SYNTHETIC, POLYPROPYLENE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key22059648
Report Number2210968-2025-05638
Device Sequence Number7634791
Product Code GAW
UDI-Device Identifier10705031232778
UDI-Public(01)10705031232778(17)290430(10)1015L4
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K133356
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberHS6857H
Device Lot Number1015L4
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2025
Supplement Date Received by Manufacturer05/23/2025
06/23/2025
Initial Report FDA Received Date05/21/2025
Supplement Report FDA Received Date06/19/2025
07/18/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-