Catalog Number BOST3211
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).B3: date of event unknown / not provided.
 
Event or Problem Description
It was reported that implant was removed due to peri-implantitis.Tooth site # 12.Symptoms as a result of the event: inflammation.
 
Additional Manufacturer Narrative
The reported device was not returned.Device malfunction could not be verified.Peri-implantitis is an infectious disease that causes inflammation of the gum and the bone structure around a dental implant.Chronic inflammation causes bone loss, which can lead to implant failure thus its removal.Investigations performed for hundreds of events where either of these conditions, infection, or bone loss, or both, have been reported, have concluded that the likely cause(s) for the implant failure in relation to these conditions are external factors.Those include medical conditions (e.G., diabetes, bruxism, etc.), patient habits (e.G., smoking, bad oral hygiene routine) and user error (e.G., surgical technique).There has not been any instance where either infection and/or bone loss, and when right conditions prevail and led to peri-implantitis whether reported or not along with the other two, have resulted from a manufacturing or design defect.Furthermore, the probability of manufacturing or design defects that might lead to infection and result in bone loss occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.The analysis and result of investigations and the analysis of probability previously described are contained in the summary investigation reports performed for bone loss and infection, which are attached.Additionally, all device history record reviews verified that each implant was sterilized per procedure for every device.All complaint data used for the summary investigation was found to be conforming and did not meet capa/hhe/d/ or any further escalations.Therefore, there were no complaints which confirmed a manufacturing or design related issue that did or could cause or contribute to the reported event.Dhr review was completed for the subject lot number 2022011151.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record (op0210) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 2022011151, for similar events and no other complaints were identified.Review completed utilizing keywords: ¿functional : peri-implantitis¿ as documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection, bone loss, and/or peri-implantitis problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.G3: date received by manufacturer.G6: checked "follow-up".H6: entered evaluation codes.H10: added manufacturer narrative.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
Peri-implantitis is an infectious disease that causes inflammation of the gum and the bone structure around a dental implant.Chronic inflammation causes bone loss, which can lead to implant failure thus its removal.Investigations performed for hundreds of events where either of these conditions, infection, or bone loss, or both, have been reported, have concluded that the likely cause(s) for the implant failure in relation to these conditions are external factors.Those include medical conditions (e.G., diabetes, bruxism, etc.), patient habits (e.G., smoking, bad oral hygiene routine) and user error (e.G., surgical technique).There has not been any instance where either infection and/or bone loss, and when right conditions prevail and led to peri-implantitis whether reported or not along with the other two, have resulted from a manufacturing or design defect.Furthermore, the probability of manufacturing or design defects that might lead to infection and result in bone loss occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming product.The analysis and result of investigations and the analysis of probability previously described are contained in the summary investigation reports performed for bone loss and infection, which are attached.Additionally, all device history record reviews verified that each implant was sterilized per procedure for every device.All complaint data used for the summary investigation was found to be conforming and did not meet capa/hhe/d/ or any further escalations.Therefore, there were no complaints which confirmed a manufacturing or design related issue that did or could cause or contribute to the reported event.Dhr review was completed for the subject lot number 2022011151.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record (b)(4) was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 2022011151, for similar events and no other complaints were identified.Review completed utilizing keywords: ¿functional : peri-implantitis¿.As documented in the summary investigations, contributing factors for the reported event likely exist outside of zimvie control, including those related to patient biological factors/condition and surgical technique.Based on the summary investigations, no malfunction occurred upon investigation.The reported events remain non-verifiable as they are a medical condition.Monthly trending to date has not identified any statistical triggers suggesting potential design or manufacturing related issues.Previously completed investigations for infection, bone loss, and/or peri-implantitis problems reported in association with implant failure have not identified any signals indicating potential non-conformances impacting the manufacturing and sterilization processes.Therefore, escalation to capa is not required.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.D9: device availability.G3: date received by manufacturer.G6: checked "follow-up".H6: entered evaluation codes.H10: added manufacturer narrative.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3I T3 NON-PLATFORM SWITCHED TAPERED IMPLANT 3.25 X 11.5MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
carlos gordian-arroyo
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key22060301
Report Number0001038806-2025-01142
Device Sequence Number12844264
Product Code DZE
UDI-Device Identifier00844868028888
UDI-Public(01)00844868028888(17)270223(10)2022011151
Combination Product (Y/N)N
Initial Reporter CountryLE
PMA/510(K) Number
K122300
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Catalogue NumberBOST3211
Device Lot Number2022011151
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date05/21/2025
Supplement Report FDA Received Date07/22/2025
07/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age43 YR
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-