|
|
| Model Number |
PMAC71RSC |
| Medical Device Problem Code |
Erratic or Intermittent Display (1182)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
According to the reporter, the cable had intermittent signal acquisition.The cable would have a signal and then an error would come up to check the sensor.Sometimes, the number would be in the "low teens ," but it would spin for a baseline.It would then give a baseline and when the sensors were checked, it did not provide a number anymore.The monitor, preamplifier, sensor, and cables were swapped, and the issues were isolated to the cable.There was no injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the cable of the unit was defective.It was reported that the monitor provided erratic oximetry readings.The reported issue was confirmed.The most likely cause was traced to component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.This report is being submitted as part of remediation activities associated with capa # 643143, which is addressing mdr reporting gu idance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs ; this is not a new malfunction/event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|