Model Number K-LESS T-ROPE IMP ONLY, SYN REPR, SS
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/30/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
On (b)(6) 2025, it was reported by a sales representative via (b)(4) that an ar-8924ss knotless tightrope implant ruptures the loop traction suture from the base of the button so they fail to reduce it.This was discovered during use in a procedure with no patient harm.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.The most likely cause for the reported failure can be attributed to user error of the device due to excessive force used during suture passing/tightening /tensioning.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
K-LESS T-ROPE IMP ONLY, SYN REPR, SS
Common Device Name
WASHER, BOLT, NUT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22061947
Report Number1220246-2025-02146
Device Sequence Number8478713
Product Code HTN
UDI-Device Identifier00888867265592
UDI-Public00888867265592
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
K043248
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 10/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberK-LESS T-ROPE IMP ONLY, SYN REPR, SS
Device Catalogue NumberAR-8924SS
Device Lot Number25236
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/30/2025
Supplement Date Received by Manufacturer10/02/2025
Initial Report FDA Received Date05/21/2025
Supplement Report FDA Received Date10/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-