• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. LGC TIBIAL FIT TRAY CEM SZ 2F / 1T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. LGC TIBIAL FIT TRAY CEM SZ 2F / 1T; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-45-2010
Medical Device Problem Code Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/11/2023
Type of Reportable Event Serious Injury
Event or Problem Description
As reported, approximately 11 months and 29 days post the initial left total knee arthroplasty, in light of unforeseen complications associated with the knee prosthesis, the surgeon deemed it necessary to perform revision surgery.It was identified that there was mobilization of the tibia, accompanied by abnormal wear of the polyethylene component.The patient underwent a total knee revision and a competitor's prosthesis was implanted.There was no reported breakage of a device or surgical delay/prolongation.
 
Additional Manufacturer Narrative
H3: the revision reported was likely the result of prosthesis wear and an insufficient bond between the tibial tray and the bone, which led to aseptic (non-infected) tibial loosening.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.The extent and root cause of the prosthesis wear and tibial loosening could not be determined as the devices were not returned for evaluation, and images were not provided.D10: (b)(6) - 200-02-29 - three peg patella 29mm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGC TIBIAL FIT TRAY CEM SZ 2F / 1T
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
3523771140
MDR Report Key22062334
Report Number1038671-2025-02109
Device Sequence Number20742918
Product Code JWH
UDI-Device Identifier10885862003898
UDI-Public10885862003898
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K101981
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2023
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number02-012-45-2010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/23/2024
Initial Report FDA Received Date05/21/2025
Date Device Manufactured10/05/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11
Outcome Attributed to Adverse Event Hospitalization;
Patient Age63 YR
Patient SexFemale
-
-