• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JIANGYIN SINBON ELECTRONICS CO LTD INVOS; OXIMETER, TISSUE SATURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JIANGYIN SINBON ELECTRONICS CO LTD INVOS; OXIMETER, TISSUE SATURATION Back to Search Results
Model Number PMAC71RSC
Medical Device Problem Codes Failure to Analyze Signal (1539); Communication or Transmission Problem (2896); Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/21/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
D10 concomitant product: pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#2023-04-28); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#2023-04-28); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#2023-04-28); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#2023-04-28); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#2023-04-28); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#2023-04-28); pmac71rsc, sensor cable re-usable pmac71rsc invos (lot#2023-04-28) h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the cable of the unit was defective.It was reported that the product was defective upon receipt, the monitor was powered on but not providing oximetry readings and the sensor was not recognized by the monitor.The reported issues were confirmed.The most likely cause was traced to component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.This report is being submitted as part of remediation activities associated with capa # 643143, which is addressing mdr reporting guidance from fdas 1995 preamble, which ensures complaints related to corrections and recalls which were previously reported to fda under 21 cfr 806 are now reported as mdrs ; this is not a new malfunction/event.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, prior to use, the cable did not work out of the box during installation.The cable was issued on 1.0 software instead of 2.After checking all monitors and pre-amps, it was noted that the lot was part of affected products and did not work out of the box.When the cable was attached, there were no readings and gave an error that stated to replace sensor.The sensor was replaced, the pre amp-was swapped, and the cable was replaced.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INVOS
Common Device Name
OXIMETER, TISSUE SATURATION
Manufacturer (Section D)
JIANGYIN SINBON ELECTRONICS CO LTD
no. 288 middle chengjiang rd.
jiangyin,100 21443 4
CH  214434
Manufacturer (Section G)
JIANGYIN SINBON ELECTRONICS CO LTD
no. 288 middle chengjiang rd.
jiangyin,100 21443 4
CH   214434
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key22062453
Report Number2936999-2025-00314
Device Sequence Number18259937
Product Code MUD
UDI-Device Identifier10884521780125
UDI-Public10884521780125
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K182868
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPMAC71RSC
Device Catalogue NumberPMAC71RSC
Device Lot Number2023-04-28
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/09/2024
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/22/2023
Initial Report FDA Received Date05/21/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-