|
MENTOR TEXAS MENTOR SMOOTH ROUND MODERATE PLUS PROFILE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Catalog Number |
3502425 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a 425cc mentor smooth round moderate plus profile saline breast prosthesis being used in a primary breast augmentation had a defected valve during the operation.No adverse event was experienced by the patient.The surgeon used backup devices, and there were no reported patient consequences or procedural delays.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: n/a.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On june 04, 2025, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On july 03, 2025, the evaluation for the device was completed.Device evaluation summary: visual analysis of the returned device revealed that no damage or anomalies were observed on the breast implant.Leak testing was performed, in accordance with mentor procedures, and no leak sites were detected during the analysis.In addition no leaks from the valve were found.The event described could not be confirmed as the breast implant was returned without detectable damage.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|