|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Product complaint # (b)(4).Investigation summary : according to the information received, "during the procedure, when inserting the trial insert, the doctor levers it out to insert a larger augmentation, and this is where the insert fractures." the product was not returned to medtech orthopaedics, however photos were provided for review.The photo investigation revealed that ¿254500504, attune fb ps artic surf sz4¿ had broken and its spring was missing.The failure mode is consistent with inserting an extractor device in between the trial and mating shim, and using the extractor to pry the mating devices apart during extraction of the trials.This improper technique results in damage to the spring and/or post components of the articulating surface as well as the mating shim.The attune intuition surgical technique 0612-10-512 (page 53), emphasizes the correct use of the tibial trial extractor (product code 254500138) with the trials.Furthermore, on page 51 of the surgical technique and per ifu-0902-00-836, careful inspection of the trials for damage/breakage should be performed pre and post-operative.If any damage to the balseal components is observed the trail should be replaced.In the event of instrument breakage during use, ensure that all device fragments that may have entered the surgical site are removed prior to completion of the procedure, as patient injury may result.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the attune fb ps artic surf sz4 would contribute to the complained device issue.Based on the investigation findings, potential cause can be attributed to unintended use error and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|